Illuminance Meter Specification Range

Apply your instrument’s additive reading-percentage and display-count specification to an illuminance reading.

Apply an instrument specification written as ±(% of reading + counts × resolution). Enter the actual instrument, range and conditions; the example is invented.

Results update as you edit.

Calculating needs JavaScript. The form becomes available when its script is ready; the guide below is always readable.

Calculation inputs

Calculation inputs
Finite and zero or more, in the same range as the specification.
Enter the percentage term for this range, for example 5 means 5%, not 0.05.
A whole number of zero or more, as defined by the instrument specification.
Positive lux per least significant display count on this range. Convert other units first.
Record the model, additive formula, resolution, source, calibration and applicable conditions. Do not substitute a percentage of full scale.

Continue your task

Keep the specification band as an instrument allowance; do not interpret it as a confidence interval or complete uncertainty budget.

If this answers your question within the model’s limits, record the original inputs, unrounded result and conditions. Otherwise choose an action only when its condition applies:

See this task in the decision tree

Opening another tool does not transfer or store your entries. Keep the original values, units and conditions with the next calculation.

Read what the percentage refers to

A specification written as ±(p% of reading + n counts) has two separate terms. The percentage grows with the reading. The count term depends on the resolution of the selected display range, so it can remain significant when the reading is small. This calculator applies that particular additive form; a percentage of full scale, a calibration correction or a differently written specification needs its own interpretation.

For reading R in lx and resolution q in lx per count:

Allowance a = R × p/100 + n × q.

The displayed specification band is max(0, R − a) to R + a, in lx. Clipping the lower end at zero reflects the nonnegative physical quantity, without turning a negative mathematical endpoint into a claim of better accuracy. The inputs still need to describe the applicable instrument range and conditions.

An artificial range at two light levels

Suppose an invented meter range has a 5% reading term, two additional counts and 1 lx per count. These are arithmetic examples, not a specification for an actual instrument.

Reading Percentage term Count term Specification band
500 lx 25 lx 2 lx 473.00–527.00 lx
10 lx 0.5 lx 2 lx 7.50–12.50 lx
0 lx 0 lx 2 lx 0.00–2.00 lx

The fixed count term is a much larger share of the low reading. A different range may change both its resolution and its published tolerance. Do not carry the 1 lx resolution into another range simply because the instrument name is the same. Results use two decimal places, with a below-display-precision marker for tiny nonzero values; calculations use unrounded inputs.

A tolerance band is not a confidence interval

NIST’s Type B uncertainty guidance explains how specifications and calibration information can contribute to uncertainty evaluation, including the role of distributions. Adding the two specification terms here does not perform that evaluation and does not establish a 95% probability.

Cosine response, spectral mismatch, calibration, drift, sensor positioning, shadows and changing light can affect a measurement. Whether any of these are already included in a particular published specification depends on that document. Keep its conditions with the result rather than adding another percentage without understanding its basis.

When interpreting sample statistics, a larger number of readings does not automatically cancel a shared meter bias. A paired daylight factor also depends on two readings and their common conditions; this single-reading band does not propagate their uncertainty through the ratio.

Related tools

Paired Daylight Factor

Calculate the indoor-to-outdoor horizontal illuminance ratio under declared overcast daylight measurement conditions.

Inputs and results
Inputs
Simultaneous indoor and unobstructed outdoor lx, with sky, plane and artificial-light conditions.
Results
Daylight-factor percentage and a warning for indoor readings above the outdoor reference; no target verdict.

Illuminance Unit Converter

Convert the same illuminance between lux, kilolux and foot-candles, with metric units shown first.

Inputs and results
Inputs
A nonnegative illuminance and its unit: lx, klx or fc.
Results
The same illuminance in lx, klx and fc.

Illuminance Sample Statistics

Summarise spatial lux or PPFD samples with a supplied reference, or describe daylight occupied-time coverage with missing periods kept explicit.

Inputs and results
Inputs
Same-basis spatial lux/PPFD and represented areas, or daylight-only zone/time records, occupied horizon and user threshold.
Results
Spatial descriptive statistics and reference shares, or daylight area-time coverage bounds, observed mean and recorded exposure.

Browse all lighting tools